Chemistry analyzer A
Glucose · calibration verification · Chemistry analyzer A
Glucose (mg/dL) · Calibration verification
Criteria met
2 of 2 questions judged · Accuracy and Range coverage met
- Instrument
- Chemistry analyzer A (CA-41027)
- Study date
- Sep 15, 2026
Table 1. Acceptance criteria
| Question | Limit | Finding | Result | |
|---|---|---|---|---|
| Accuracy | Each mean within its bias limit; each replicate within its replicate limit | 5 of 5 materials within | Met | |
| Range coverage | Low and high within 10% of 2 and 750 mg/dL (or zero at the low end), with a middle material | Covered | Met |
Table 2. Accuracy by material
| Level | Assigned (mg/dL) | Mean (mg/dL) | Bias (mg/dL) | Bias limit (mg/dL) | Replicate limit (mg/dL) | Within / total | Accuracy |
|---|---|---|---|---|---|---|---|
| 1 | 2 | 2.3 | 0.3 | 3.0 | 6.0 | 3/3 within | |
| 2 | 189 | 188.3 | −0.7 | 7.6 | 15.1 | 3/3 within | |
| 3 | 376 | 376.3 | 0.3 | 15.0 | 30.1 | 3/3 within | |
| 4 | 563 | 562.7 | −0.3 | 22.5 | 45.0 | 3/3 within | |
| 5 | 750 | 747.7 | −2.3 | 30.0 | 60.0 | 3/3 within |
Study details
- Matrix
- Serum
- Method
- Hexokinase
- Reagent lot
- GLU-R2609
- Calibrator lot
- GLU-C2609
- Data
- 15 included
- Design
- 5 material levels
A Data listing
Source: CHEM-A_LIN-2609_20260915.csv. 15 rows, all included. Every row: within total-error allowance.
| Study row | Material | Level | Result (mg/dL) | Replicate | Source row |
|---|---|---|---|---|---|
| 1 | LIN-2609 L1 | 1 | 2 | 1 | 2 |
| 2 | LIN-2609 L1 | 1 | 3 | 2 | 3 |
| 3 | LIN-2609 L1 | 1 | 2 | 3 | 4 |
| 4 | LIN-2609 L2 | 2 | 188 | 1 | 5 |
| 5 | LIN-2609 L2 | 2 | 190 | 2 | 6 |
| 6 | LIN-2609 L2 | 2 | 187 | 3 | 7 |
| 7 | LIN-2609 L3 | 3 | 378 | 1 | 8 |
| 8 | LIN-2609 L3 | 3 | 374 | 2 | 9 |
| 9 | LIN-2609 L3 | 3 | 377 | 3 | 10 |
| 10 | LIN-2609 L4 | 4 | 560 | 1 | 11 |
| 11 | LIN-2609 L4 | 4 | 566 | 2 | 12 |
| 12 | LIN-2609 L4 | 4 | 562 | 3 | 13 |
| 13 | LIN-2609 L5 | 5 | 744 | 1 | 14 |
| 14 | LIN-2609 L5 | 5 | 752 | 2 | 15 |
| 15 | LIN-2609 L5 | 5 | 747 | 3 | 16 |
1 Results
1.1 Supporting estimates
Table 3. Intervals and error allowances
| Level | Mean 95% CI (mg/dL) | Bias (%) | Recovery (%) | SD (mg/dL) |
|---|---|---|---|---|
| 1 | 0.9–3.8 | 16.7 | 116.7 | 0.577 |
| 2 | 184.5–192.1 | −0.4 | 99.6 | 1.53 |
| 3 | 371.2–381.5 | 0.1 | 100.1 | 2.08 |
| 4 | 555.1–570.3 | −0.1 | 99.9 | 3.06 |
| 5 | 737.6–757.7 | −0.3 | 99.7 | 4.04 |
1.2 Range coverage · Middle material: 3.
Table 4. Endpoint coverage
| Endpoint | Window (mg/dL) | Material | Result |
|---|---|---|---|
| Lower | 1.8–2.2 | 1 | Met |
| Upper | 675.0–825.0 | 5 | Met |
B Methods
B.1 Additional study details
| CAP COM.40300 · 9 December 2025 | Unmodified FDA-cleared or approved nonwaived tests. Supporting sections: Decision page; Results; Data listing. Associated laboratory evidence: Written assessment for each test and instrument, including investigation of discordant results. |
| CAP CHM.13550 · 9 December 2025 | Chemistry AMR verification materials covering low, middle and high concentrations. Supporting sections: Range coverage; Data listing. Associated laboratory evidence: Material suitability and preparation instructions. |
| CAP CHM.13600 · 9 December 2025 | Chemistry AMR verification with applicable timing, triggers and exceptions. Supporting sections: Range coverage; Decision page. Associated laboratory evidence: AMR schedule and applicable event records. |
| CAP COM.40475 · 9 December 2025 | Approval of nonwaived test validation or verification before clinical use. Supporting sections: Decision page; Sign-off. Associated laboratory evidence: Director or qualified designee signed written assessment and approval tied to each test and instrument. |
| CAP GEN.20377 · 9 December 2025 | Laboratory record retention. Supporting sections: Evidence archive. Associated laboratory evidence: Keep method validation and verification records for the life of the test plus two years, or longer when applicable retention rules require it. |
| CAP GEN.20450 · 9 December 2025 | Corrections to laboratory records. Supporting sections: Data listing; Audit trail. Associated laboratory evidence: Retained original values and traceable corrections. |
B.2 Statistical methods applied
- Independent assigned values or target ranges determine accuracy; measured results never assign their own accuracy targets.
- Total-error allowance is the larger of the entered concentration allowance and percentage of the assigned value. Systematic and random allowances are their entered shares of total error.
- Mean bias and every replicate are compared directly with inclusive laboratory limits. A confidence interval does not relax a typed limit.
- Assigned-high dilution uses percent × assigned high / 100. Two assigned anchors use linear interpolation in the entered percentages or level indices.
- Target-range midpoints locate materials for coverage only; no mean bias is calculated from a range midpoint.
- Range coverage needs a material covering each end, a distinct interior material, and acceptable accuracy at every planned level. A material covers an end inside its endpoint window, limit ± max(concentration allowance, percentage × |limit| / 100), with 10% for a blank allowance; a minimal or zero material from zero up to the lower limit also covers the lower end. Of the covering materials, the one nearest the limit covers that end.
- Means and sample SDs summarize within-run replicates. Mean confidence intervals use Student t and n − 1 degrees of freedom.
B.3 Decision rules and display precision
- Each acceptance criterion compares the unrounded estimate with its limit. Limits include their end values unless marked exclusive. A criterion with a missing estimate, too little data or a failed fit is inconclusive and does not count as met.
- Entered results, claims, targets and acceptance limits are shown as entered. Calculated concentrations, SDs and fitted parameters are shown to 1 decimal place, percentages to 1 decimal place, and slopes to 3 decimal places. Very small or large values use exponent notation at the same precision. An SD or limit that would round to zero keeps two significant digits.
C Audit trail
| Data version | 3 |
| Criteria version | 2 |
| Method | calibration verification (method ID LMS-CAL-1.3) |
| Criteria evaluation | acceptance criteria evaluation (method ID LMS-CRITERIA-1.1) |
| Import CHEM-A_LIN-2609_20260915.csv | SHA-256 30af7040d07cc68002d413138d5d2612510f2201ee0f85ab7164e8d77762c87d |
| Report software | 0.1.0 · templates 2a6731123b96 |